바로가기 메뉴
컨텐츠바로가기
주메뉴바로가기
하단메뉴바로가기

PRODUCTS

Industrial Test Equipment

  • HOME
  • > PRODUCTS
  • > Industrial Test Equipment
Industrial Test Equipment

Heating & Cooling Chambers (PBV)

Temperature values of all chambers are measured according to DIN 12880 and IEC 60068-3-5.

Model Name : PBV-012/025/040/070/100


P - SERIES

Two-point door latch

Stainless steel chamber

Perforated shelves for heavy load (optional)

Color LCD controller

Automatic start

Program control

End of program

Function setup

Operating status

Program setup

Graph display

Performance

  • Suitable for specific use such as heat treatment and drying process.
  • Microprocessor PID control / Temperature calibration / Automatic tuning.
  • Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
  • Two-point door latch also enhances tightness to door closure.
  • Energy efficiency is achieved thanks to optimization of the insulation system.
    - Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 51℃. (DIN EN 563)
  • Reliable design and production of door is guaranteed.
    - Evaluates the results after opening the door by 90 degrees and applying 50kg load to the point of 30cm from hinges.

Safety

  • User selectable operation mode after power interruption
    - ether Automatic run or Manual run.
  • Over-temperature protection.
  • Over-current and current leakage protection. (ELCB, earth leakage circuit breaker)
  • Independent safety fuse.
    - Important parts such as controller and heaters are highly secured from short circuit. Especially each heater has an individual fuse to completely break a short circuit.
  • Reversed-phase protection.
    - Electrical connection error indicator helps user to rapidly fix the connection.
  • Over-current protection of motor.
    - Automatic shut-off in situations where there is over-current to the motor.
  • Open door warning.
    - Heater and fan stop when the door is the open state.
  • Electric wiring panel access warning.
    - Automatic stop if the cover of electric wiring panel is opened.
  • Built in door lock with a key.

Convenience

  • Corrosion resistance stainless steel inner chamber.
  • Space efficient and easy loading chamber design.
    - Included two shelves for loading of large amounts of samples.
    - Heavy load shelving, rated for 40kg are also available. (optional)
  • User friendly LCD controller.
    - Colorful interactive touch type LCD controller.
    - Set point temperature and present temperature are clearly shown together on display.
    - Sufficient programmable control support.
    - Operating progress can be shown in graph mode for better data understanding.
    - Automatic start/stop can be programmed for unattended operations.
    - Display indicates abnormal temperature changes or power failures.
    - Three different language support. (English / Chinese / Korean)
    - Temperature and other operating parameters can be recorded using a six channel digital recorder. (optional)
  • Computer interface.
    - RS-485 communication port (optional) enables connection up to maximum 32 units. Simultaneous monitoring and operating of multiple units is an available feature.
    - Program patterns are simple to set with the use of our software.
    - Easy to check status of real-time operations. Graph data supports easy understanding.
    - Storage and documentation of programs.
    - Data with saving capabilities in excel and text format.
    - Storage of program information, backup of value settings, recovery, and temperature values.
    - RS-232 communication port as standard offer
Description No
Programmable pattern capacity 100
Pattern repeat time 999
Part repeat time 255
Max. segments per a pattern 100
Available max. segments* 2000
Programmable process time per a segment 0 to 99 hr. 59 min.

* Even though the max. segments per pattern are 100, the total available segments are up to 6000 not 10000, therefore arrangement of the number of segment and patterns will need to be done properly.

       
Model PBV-012 PBV-025 PBV-040 PBV-070 PBV-100
Temperature data
Range (℃ / ℉) -25 to 100
/ -13 to 212
-25 to 100
/ -13 to 212
-25 to 100
/ -13 to 212
-25 to 100
/ -13 to 212
-25 to 100
/ -13 to 212
Fluctuation (±℃ / ±℉) 1), 2)
at -25℃ 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36
at -15℃ 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36
at -10℃ 0.3 / 0.54 0.3 / 0.54 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36
at 25℃ 0.3 / 0.54 0.3 / 0.54 0.3 / 0.54 0.4 / 0.72 0.2 / 0.36
at 70℃ 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.3 / 0.54 0.3 / 0.54
at 100℃ 0.2 / 0.36 0.2 / 0.36 0.2 / 0.36 0.3 / 0.54 0.2 / 0.36
Gradient (±℃ / ±℉)1)
at -25℃ 0.3 / 0.54 0.3 / 0.54 0.4 / 0.72 0.4 / 0.72 0.4 / 0.72
at -15℃ 0.3 / 0.54 0.3 / 0.54 0.4 / 0.72 0.4 / 0.72 0.4 / 0.72
at -10℃ 0.3 / 0.54 0.3 / 0.54 0.4 / 0.72 0.4 / 0.72 0.4 / 0.72
at 25℃ 0.3 / 0.54 0.3 / 0.54 0.4 / 0.72 0.4 / 0.72 0.4 / 0.72
at 70℃ 0.4 / 0.72 0.4 / 0.72 0.6 / 1.08 0.6 / 1.08 0.6 / 1.08
at 100℃ 0.6 / 1.08 0.6 / 1.08 0.7 / 1.26 0.8 / 1.44 0.8 / 1.44
Heating time (min)2)
from -25℃ to 25℃ 19 19 29 22 24
from -15℃ to 25℃ 17 17 22 20 22
from -10℃ to 25℃ 15 15 21 18 20
from 25℃ to 70℃ 20 20 33 22 27
from 25℃ to 100℃ 34 34 52 35 38
from -25℃ to 100℃ 47 47 76 46 61
Cooling time (min)2)
from 25℃ to -25℃ 36 36 26 25 34
from 25℃ to -15℃ 21 21 21 20 24
from 25℃ to -10℃ 17 17 19 18 21
from 70℃ to 25℃ 17 17 20 20 25
from 100℃ to 25℃ 28 28 30 29 35
from 100℃ to -25℃ 65 65 50 53 56
Recovery time after door was opened for 30sec. (min)2)
at -25℃ 5 5 5 4 7
at -15℃ 4.5 4.5 3.5 5.5 5
at -10℃ 6 6 4.5 4 5
at 25℃ 5 5 4 5.5 3.5
at 70℃ 5 5 9.5 5 7.5
at 100℃ 5.5 5.5 9 6 7
Electrical requirement (230V, 60Hz, 1Ph) 16.3 22.8
Cat. No. AAHK5011K AAHK5021K
Electrical requirement (230V, 50Hz, 1Ph) 16.3 22.8
Cat. No. AAHK5015K AAHK5025K
Electrical requirement (380V, 50/60Hz, 3Ph) 10.0 15.4 16.8
Cat. No. AAHK5030K AAHK5040K AAHK5050K

1) 以IEC 60068-3-5为标准。
2) 以DIN12880为标准。
※ 以上规格值以230V/60Hz, 380V/50Hz记录。
※ 以上规格如有变更,概不另行通知。

  • Digital recorder

  • Perforated shelf

  • Warning signal lamp



for Floor Standing Chambers

Cat. No. Description Models
PBV/PMV
AAA8T505

Recorder (digital, 6 channels)

AAA8T540

Interface converter (RS - 485 → 232) included a cable (5m / 16.4ft)

AAA80550

Warning signal lamp

00LTV0000070

Wire shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-012 (462x470mm / 18.2x18.5”)

00LTV0000053

Wire shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-025 (562x620mm / 22.1x22.4”)

00LTV0000029

Wire shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-040 (712x670mm / (28.0x26.4”)

00LTV0000030

Wire shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-070 (862x770mm / 33.9x30.3”)

00LTV0000057

Wire shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-100 (962x880 / 37.9x34.6”)

AAAK1501

Perforated shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-012 (462x470mm / 18.2x18.5”) (max. load 40kg)

AAAK1502

Perforated shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-025(562x620mm / 22.1x22.4”) (max. load 40kg)

AAAK1503

Perforated shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-040 (712x670mm / (28.0x26.4”) (max. load 40kg)

AAAK1504

Perforated shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-070 (862x770mm / 33.9x30.3”) (max. load 40kg)

AAAK1505

Perforated shelf for LBV/LTV/KBD/KMV/JMV/PBV/PMV-100 (962x880 / 37.9x34.6”) (max. load 40kg)

       

       

Related Products

Inquiry Form

  • Title (*)

    Position

  • First name (*)

    Last name (*)

  • Company (*)

  • Business type (*)

    Business division

  • Country

    City

  • E-Mail (*)

    Phone

  • Request Product (Model)

  • Your message

  Consent to use of data *
I consent to the processing of the data above for service communication with JEIO TECH.
I agree that JEIO TECH may collect, store, and use the data that I have provided.